1-4 General Information 6222 User’s Manual
Synchronization
To synchronize from two to nine 6000 Series devices, link
them together with synchronization cables as indicated in the
example figure. Note that the first device (relative to the
circuit) will be the Master. The last device connected will be
the Terminating Slave. Thus the synchronized devices must
proceed sequentially from the Master to Terminating Slave, as
indicated in the figures. Note that channel-to-channel phase
relationships in a multi-device configuration are not
necessarily fixed. See notes below.
After physically linking the devices via the sync cables, they must still be configured as a “Synchronized Device
Group” in the Encore software. To do this, use the Synchronize Devices page of Encore‟s Hardware Setup tab.
Each Synchronized Group consists of one Master Device
and at least one Slave Device. By default, the first 6000
Series device added to the Synchronization Group will be
the Master and the last device added to the group will be the
Terminating Slave.
Encore‟s Synchronized Device Group Configuration
panel includes a pull-down list which allows you to select
the device that you want to have function as Master and the
one you want to function as Terminating Slave, should this
be different than the default settings.
Encore will display a Synchronization Error icon if it
detects that a necessary synch cable is missing or not
properly connected.
Up to nine units can be synchronized. The total combined length of the SYNC cables is not
to exceed 2.438m (8 ft).
Unplug sync cables from devices that are not intended to be part of a “Synchronized Device
Group.” Leaving a sync cable connected to an independent device can lead to errors.
Leaving an un-terminated sync cable on a synchronized device can lead to errors.
The sampled data phase relationship among channels between multiple devices is
dependent on the Channel Sync Skew specification for each device. When utilizing
different models in a multi-device system, any differences in the ADC filter delay
(“Input Delay”) should be added.
Reference Note: For detailed information refer to the Encore software‟s Comprehensive Help.
You can access the Help from within Encore. After doing so, navigate to the synchronization
related topic as follows from the Help‟s Contents panel:
Configuring Hardware>>Hardware Setup>>Synchronizing Devices.
TO ACQ
TO ACQ
TO ACQ
TO ACQ